16 µl BigDye Terminator Ready Reaction Mix 1 µg BAC DNA (determined from gel) 10 pmol primer water to 40 µl Heat tubes at 95℃ for 5 min., then perform 30 cycles of: 95℃ x 30 sec 55℃ x 10 sec 60℃ x 4 min Run BACs on a 377 and load the entire sample after clean up. If you use a 3700, the loading would be very different. This IS the double recipe. Normally a reaction is only 20 µl. For sequencing we use ABIs "BigDye Terminator Cycle Sequencing Ready Reaction Kits v2.0 with AmpliTaq DNA Polymerase", cat. no. 4314414, 100 reactions. Avoid repeated freeze-thaw of ready reaction mix. Signal decreases with each thawing. To get a protocol order cat. no 4303237 (they are free, but must be requested). Check with the group running the sequencer. Some 3700 users now are using the new BigDye 3 system. This is not available for 377 yet (different dyes require different filters). Be sure they know exactly what type of kit you used so they run the right detection and analysis parameters. For clean up we use Princeton Separations (732 431-3338) CentriSep Columns, cat. no. CS-901, 100 columns. |
→如果您认为本词条还有待完善,请 编辑词条
上一篇Non-radioactive?Probes 下一篇Extraction and PCR Analyis of DNA from Mice Tails
词条内容仅供参考,如果您需要解决具体问题
(尤其在法律、医学等领域),建议您咨询相关领域专业人士。
0