生命经纬知识库 >>所属分类 >> 常用实验技术   

Maxam-Gilbert Sequencing

标签: Maxam-Gilbert Sequencing 测序法

顶[0] 发表评论(31) 编辑词条

A modified, abbreviated version of the Maxam-Gilbert sequencing is described. A complete description including the methods to prepare and isolate asymmetrically labeled fragments of DNA and to prepare and run sequencing gels can be found in Methods in Enzymology 65: 497-559.

Reagents:

-Dimethylsulfate (DMS) Extremely dangerous.

-Hydrazine (HZ) Very dangerous.

-Formic acid

-Piperidine. Stock is 10 M. Dilute to 1.0 M just before use.

-100% Ethanol

-dH2O

-Acetic Acid. Glacial is 17.4 M. Dilute to 1 M.

-0.3 M Sodium acetate (pH 5.2)

-5 M NaCl

-10 N NaOH

-500 mM EDTA

-tRNA. Stock solution is 1mg/ml in dH2O.

-Calf thymus DNA. Stock solution is 1 mg/ml in TE.

Buffers:

DMS buffer

50 mM sodium cacodylate (pH 8.0)

1 mM EDTA

DMS stop

1.5 M sodium acetate (pH 7.0)

1.0 M mercaptoethanol

100 μg/ml tRNA

HZ Stop

0.3 M sodium acetate

0.1 M EDTA

25 μg/ml tRNA

Loading buffer

80% (v/v) formamide

50 mM Tris-borate (pH 8.3)

1 mM EDTA

0.1% (w/v) xylene cyanol

0.1% (w/v) bromophenol blue.

Hints:

1. 200 cps per 5 μl reaction or greater is recommended.

2. -70℃ incubations can be done on powdered dry ice.

3. 30 minute 90℃ incubation should be done with a device which holds Eppendorfs closed.

4. During multiple ethanol precipitations monitor presence of pellet with Geiger counter.

Nucleotide sequencing gels:

All gels are pre-run at 1200 volts for 30' (do not go longer on 20% gels)

Nucleotides 1-50

Use a 20% gel,

Run BPB to 15cm for nucleotide 1 on bottom. Alternatively, for a better spread, run BPB to 20cm for nucleotide 3 on bottom. To do this, run the gel at no more than 28 milliamps, 2000-2500 volts.

At 2000 volts, this takes 1 1/4 hours.

Nucleotides 40-150

Use an 8% gel.

Run XC to 27 cm. To do this, run the gel at no more than 28 milliamps. Voltage should start at

about 1500 volts and drops during run as plates heat up.

The run should take 3 - 3 1/2 hours.

Nucleotides 140-?

Use an 8% gel.

Run XC to 2 gel lengths (72 cm). Re-load dyes when XC is at 36 cm. To do this, run the gel at

no more than 28 milliamps. Voltage should start at about 1500 volts and drops during run.

The run should take 7 - 8 hours.

附件列表


→如果您认为本词条还有待完善,请 编辑词条

上一篇关于微型真空泵与气体采样泵的区别 下一篇多功能微板生物活性测试系统(POLARstar OPTIMA)使用程序(图)

词条内容仅供参考,如果您需要解决具体问题
(尤其在法律、医学等领域),建议您咨询相关领域专业人士。
0

收藏到:  

词条信息

admin
admin
超级管理员
词条创建者 发短消息   

相关词条